4OXV OXYGEN SENSORS AAY80-390

4OXV OXYGEN SENSORS AAY80-390

4OXV OXYGEN SENSORS AAY80-390

Category: Brand:

4OXV OXYGEN SENSORS

AAY80-390

Brand City Technology
Model Number AAY80-390 (4OXV)
Technology Electrochemical
Measurement Range 1-25% vol. O2
Maximum Overload 30% vol. O2
Output Signal* 0.10 ± 0.02 mA in Air
Response Time (T90) <15 Seconds
Zero Current (Offset) <0.6% vol. O2
Linearity Can be considered linear in
many cases.
See Operating Principles
(OP-02) for further details.
Recommended Load Resistor 100 Ω
Casing Material ABS
Weight <16 g
Orientation Sensitivity <0.2% vol. O2
equivalent
Operating Temperature Range -20°C to +50°C (up to 3 months
continuous across RH range)
Recommended Storage Temp 0°C to 20°C
Thermal Transient <23.5% vol. O2
Operating Pressure Range Atmospheric ± 20%
Pressure Coefficient <0.02% signal/mbar
Pressure Transient <200% signal change
Relative Humidity Range (at 0°C to 20°C)
Long Term Output Drift* <2% signal/month
Typically <5% over operating life

 